Application Note: Moisture Determination of Cement Raw Mix
Precisa moisture determination XM60/EM120 Halogen
This information is intended as a guideline for development of a moisture determination method and parameter setting. Materials vary throughout the world and we will be more than happy to test customers’ individual samples.
Description: | Cement raw mix, fine powder |
Utensils: | Aluminium pan |
Preparation: | Spread thinly on the pan |
Sample weight (+/- 10%): | 2g |
Temperature program: | Standard |
Temperature: | 105 oC |
Standby Temperature: | 100 oC |
Stop-Criteria: | Autostop 2d / 60 s |
Moisture content: (average of 4 measurements) | 0.68% |
Standard Deviation (+/-): | 0.04% |
Average analysis time: | 1.8 min |