Setting Serial port

A PC or printer can be connected via the RS232/V24 DB9 serial interface and a PC can be connected via the USB DEVICE serial interface. Both connections are located on the back of the balance. For data transmission, the interface parameters of the balance must match with those of the peripheral device.

Data reception

A list of possible remote control commands is provided at the end of this help text.

Data sending

Mode Protocol

In mode Protocol, the protocol defined for the application is output automatically or with the [PRINT]-key, depending on the setting.

Mode Measurement transfer

In mode Measurement transfer by pressing the soft button [Transfer] or the [PRINT]-key respectively (depending on the trigger setting) the current weight value is transferred according the set format of the settings Measurement transfer. If trigger is set on Transfer button the soft button [Transfer] must be activated in the Action bar.

Note, to ensure to activate on the Action bar the soft button [Transfer] if trigger is set on Transfer button.

 

Remote control commands

Each remote command has to be closed with CR LF which corresponds to pushing the [Enter] key. Commands can be acknowledged using ACK1.

Command specification

AOST       : Print out AOS-Temperature in degrees Celsius

AOSH       : Print out AOS-Humidity in %

AOSP       : Print out AOS-Pressure in hPa

ACK0       : Acknowledge sending mode Off

ACK1       : Acknowledge sending mode On

BINFO      : Print out Balance information

CAL        : Start adjustment

D......    : Write to weight display (right-aligned)

DN         : Reset weight display

@......    : Write to secondary display (right-aligned)

@N         : Reset secondary display

FWHISPRINT : Print out update history

PDT        : Print out date and time

PRT        : Start printing (corresponds to pressing the [PRINT] key)

P0         : Individually print each value (unstable or stable)

P1         : Individually print value (stable)

P2 lll     : Print after change of load (lll = min. load change necessary in g or in percent) 

             (Examples: P2 12.3 or P2 50%)

P3         : Print after each display refresh (refresh rate)

P4 ttt.t   : Print on time basis ttt.t in s, from 0.1s to 600.0s 

             (Example: P4 2.0 individual print out every 2s without waiting until stable)

SDTddmmyyhhmmss: Set date/time if Date format = ddmmyy

SDTmmddyyhhmmss: Set date/time if Date format = mmddyy

SOUND      : Beep tone

SOUNDI     : Info tone

SOUNDW     : Warn tone

T (ttt)    : Tare or set tare to a specific value

WSMOVE     : Move draft shield

Z0         : Auto-Zero mode Off

Z1         : Auto-Zero mode On

ZERO       : Zero balance (provided weight is within the zero-position range)

 

Application Counting:       

Rnnn       : Set current weight = nnn pieces (corresponds to soft button "Set Pcs")

REFrrr     : Set reference weight = rrr for 1 piece (corresponds to soft button "Set piece weight")

RC         : Clear pieces unit (corresponds to soft button "Clr Pcs")

 

Application Percent:       

R%k        : Set current weight = 100% with k=0...7 decimal places

             (k=A: use automatic positioning of decimal point)

             (corresponds to soft button "Set 100%")

REF%k rrr  : Set reference weight = rrr for 100% with k=0...7 decimal places

             (k=A: use automatic positioning of decimal point)

R%C        : Clear percentage unit (corresponds to soft button "Clr %")